|
Model Number AC-2003592 |
Device Problems
Defective Device (2588); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/28/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
D3.Manufacturer email: (b)(6).
|
|
Event Description
|
It was reported that the suction tube for the procedure was worn out by the heat of the roller pump, causing the tube to cut.The suction tube was cracked, and powder was coming out, and the suction was not working properly.This event is being reported for aborted/cancelled procedure with a patient whose sedation status was unknown.
|
|
Event Description
|
It was reported that the suction tube for the procedure was worn out by the heat of the roller pump, causing the tube to cut.The suction tube was cracked, and powder was coming out, and the suction was not working properly.This event is being reported for aborted/cancelled procedure with a patient whose sedation status was unknown.
|
|
Manufacturer Narrative
|
Upon receipt of this device at our quality assurance laboratory, this device was thoroughly analyzed.Visual analysis identified scuff marks on the surface of the tube.It is likely that the scuff marks were caused by the interaction between the tube and the pump.It is probable that the tube could have been placed incorrectly on the pump, which could have lead the heating of the pump, causing the damage to the suction tube.Instructions for use (ifu) state: do not use tubing other than that provided by lumenis.Using alternative tubing may jeopardize the safe operation of the instrument.Aspiration tubing is only qualified for single use and may not be resterilized.Discard after use.Insert the sterile aspiration tubing into the pump head, as shown.Leave sufficient tubing at the right-hand side to reach the tissue catch while allowing for maximum tubing to the handpiece.Ensure that the tubing is seated securely within the pump head, on top of the metal rollers and under the black side clamps.Aspiration flows in the direction of the arrow on the pump head.Always verify that the aspiration tube is loaded in the required direction.Use the lowest effective aspiration rate until you are comfortable and proficient with the instrument.
|
|
Search Alerts/Recalls
|
|
|