MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 7122Q/58
Device Problems Difficult to Fold, Unfold or Collapse (1254); Retraction Problem (1536); Capturing Problem (2891)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/17/2023
Event Type  Injury  
Event Description
It was reported that the patient right ventricular lead exhibited high capture threshold.The patient presented for a lead revision.During the procedure, the lead's helix had difficulty retracting and would not extend afterwards.The lead was explanted and replaced.The patient condition was stable.
 
Manufacturer Narrative
The reported event of high pacing threshold was not confirmed while the reported event of helix mechanism issue was confirmed.As received, a complete lead was returned in one piece.Visual inspection of the lead found the helix was retracted clogged with blood/tissue.The cause of the reported event of helix mechanism issue was isolated to the helix being clogged with blood/tissue.Electrical tests and x-ray examination were normal.No problem was detected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17363008
MDR Text Key319414964
Report Number2017865-2023-36327
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734503174
UDI-Public05414734503174
Combination Product (y/n)Y
Reporter Country CodeTW
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7122Q/58
Device Catalogue Number7122Q-58
Device Lot NumberA000133876
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/17/2023
Initial Date FDA Received07/20/2023
Supplement Dates Manufacturer Received08/16/2023
Supplement Dates FDA Received08/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/14/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-