Model Number N/A
Device Problems Break (1069); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2023
Event Type  Injury  
Event Description
It was reported an 81 year-old patient underwent placement of an ultrathane mac-loc locking loop multipurpose drainage catheter to address cholestasis on (b)(6) 2023 at 09:10.The drainage catheter was secured to the patient with suture.On (b)(6) 2023 at 09:50 the nurse discovered that the drainage catheter was broken 1 cm from the hub from the patient's body.The drainage catheter was replaced on (b)(6)203 at 10:12.It is suspected that the catheter may have broken when the patient turned over.The device was discarded by the hospital and will not be returned.
 
Manufacturer Narrative
E1 - customer (person): street: (b)(6).G4 - pma/510(k) #: exempt.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Correction: h6 - annex a investigation ¿ evaluation (b)(6)reviews of documentation including the complaint history, device history record (dhr), quality control procedures, specifications, and instructions for use (ifu) of the complaint device were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhrs for the reported complaint device lot 14874670 and the related subassembly lots revealed no relevant non-conformances.To date, a further search of our database records revealed no other complaints have been received associated with the reported lot.Cook also reviewed product labeling.The ifu, [ifu__multi2_rev1], packaged with the device contains the following in relation to the reported failure mode: precautions patients with indwelling drainage catheters should be evaluated routinely to ensure continuous function of the catheter.How supplied upon removal from package, inspect the product to ensure no damage has occurred the information provided upon review of the dmr and ifu suggests that the device was not manufactured out of specification, and that there are no nonconforming devices in house or out in the field.Based on the information provided, no product returned, and the results of the investigation, the cause of this event was determined to be component failure, without any design or manufacturing issue.The user did not identify any damage during initial placement.It is possible that the catheter underwent excessive force while placed in the patient, but this cannot be confirmed without additional information.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER
Type of Device
GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key17363289
MDR Text Key319418616
Report Number1820334-2023-00940
Device Sequence Number1
Product Code GBO
UDI-Device Identifier00827002095014
UDI-Public(01)00827002095014(17)250803(10)14874670
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberULT8.5-38-25-P-6S-CLM-RH
Device Lot Number14874670
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/13/2023
Initial Date FDA Received07/20/2023
Supplement Dates Manufacturer Received08/31/2023
Supplement Dates FDA Received09/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age81 YR
-
-