Model Number 87035 |
Device Problems
Detachment of Device or Device Component (2907); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/28/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been received for analysis; therefore, a failure analysis of the complaint device could not be completed.
|
|
Event Description
|
It was reported that after an ablation procedure had been completed using an intellamap orion high resolution mapping catheter (orion) they found one of the array splines was detached from the distal tip.The procedure had been completed using the original device and the damage was only found when the catheter was withdrawn.No patient complications were reported.The catheter is expected to be returned for analysis.
|
|
Manufacturer Narrative
|
The catheter was not returned to boston scientific for analysis; however, images were provided from the field.Inspection of the photos identified that a spline was detached from one end, at the distal tip of the catheter.The allegation of catheter damage from the field was confirmed.The most likely cause was determined to be an adverse event related to the procedure, as the detached spline likely occurred during use of the catheter during the procedure.
|
|
Event Description
|
It was reported that after an ablation procedure had been completed using an intellamap orion high resolution mapping catheter (orion) they found one of the array splines was detached from the distal tip.The procedure had been completed using the original device and the damage was only found when the catheter was withdrawn.No patient complications were reported.The catheter has not been received for analysis, however, images of the defective catheter were provided and reviewed by investigators.
|
|
Search Alerts/Recalls
|