This report is being supplemented to provide additional information based on the legal manufacturer's final investigation, as well as corrections to d9 and h3.Please see updates to d9, h3, h6 and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 4 years since the subject device was manufactured.Based on the results of the investigation, it¿s likely the damaged insulation at the distal end is due to improper handling by the user during reprocessing.Additionally, it¿s likely the deformation was caused by excessive force by the user.The final root cause of this event was unable to be identified.The following is included in the instructions for use: ¿warning! infection control risk: properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.Warning! risk of injury: defective electrodes are dangerous for the patient as well as for the operating surgeon and must not be used.Never attempt to repair defective hf electrodes, always replace them.Warning! risk of injury: impact, fall, shock, or similar stress can result in damage of the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries of the patient and/or user.Do not use the instrument if damaged.¿ olympus will continue to monitor field performance for this device.
|