• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM Back to Search Results
Model Number AB2000
Device Problems Corroded (1131); Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2023
Event Type  malfunction  
Manufacturer Narrative
H.10 additional manufacturer narrative: h.6 adverse event problem.Component code 4756: per the instructions for use, the aquabeam motorpack, a re-usable component of the aquabeam robotic system, provides power to the aquabeam handpiece by means of dc motors.Root cause of reported event has not yet been established.Investigation by manufacturer is currently in-process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation became aware that during procedure, an "e22 - motorpack error" message was generated by the aquabeam robotic system.A second handpiece was used to resolved the issue, but the error persisted.As a result, the aquablation procedure was aborted and converted to transurethral resection of the prostate (turp) surgery.Additionally, it was observed that some corrosion was present on the aquabeam motorpack.There were no adverse health consequences to the patient due to this event.
 
Manufacturer Narrative
The aquabeam motorpack was returned for investigation.Visual inspection revealed signs of fluid ingress in and around the connector port, which confirmed the reported event.The root cause of the reported error is fluid ingress; however, the cause of the fluid ingress is undeterminable.A review of the device history record (dhr) ab2000-b / serial number (b)(6) and aquabeam motorpack / lot number 21c01779 was conducted, which confirmed there were no non-conformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system met all design and manufacturing specificiations when released for distribution.The current user manual um0104-00 rev.G, aquabeam robotic system user manual, intl (ce), english was reviewed.Table 5 system detected errors and faults e22 - motorpack error release foot pedal and click x.1) if error persists, reconnect handpiece to motorpack.2) if error continues, replace handpiece.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
900 island drive
suite 101
redwood city CA 94065
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
150 baytech drive
suite 101
san jose CA 95134
Manufacturer Contact
doria esquivel
900 island drive
150 baytech drive
san jose, CA 95134
6502327291
MDR Report Key17366843
MDR Text Key319784472
Report Number3012977056-2023-00122
Device Sequence Number1
Product Code PZP
UDI-Device IdentifierB614AB20001
UDI-Public+B614AB20001/16D20190307P
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAB2000
Device Catalogue NumberAB2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/06/2023
Initial Date FDA Received07/20/2023
Supplement Dates Manufacturer Received10/06/2023
Supplement Dates FDA Received10/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/07/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-