Model Number MS20401 |
Device Problem
No Audible Prompt/Feedback (2282)
|
Patient Problem
Low Oxygen Saturation (2477)
|
Event Date 05/17/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A follow-up report will be submitted upon completion of this investigation.
|
|
Event Description
|
It was reported that the infinity acute care system (iacs) did not provide the expected high priority (red) alarm (red), it provided a medium priority (yellow) alarm on a neonatal patient in the neonatal intensive care unit (nicu).The spo2 was set to 90 and the patient went as low as 60.There was no patient injury.
|
|
Event Description
|
It was reported that the infinity acute care system (iacs) did not provide the expected high priority (red) alarm (red), it provided a medium priority (yellow) alarm on a neonatal patient in the neonatal intensive care unit (nicu).The spo2 was set to 90 and the patient went as low as 60.There was no patient injury.
|
|
Manufacturer Narrative
|
The concerned device was inspected by dräger on-site.It was found that the so-called "satseconds" function was enabled.The intention of this function is the suppression of nuisance alarms when thresholds are only exceeded briefly but frequently.The degree of suppression can be adjusted by the user.As per report, the user was expecting a high-priority desaturation alarm in the event situation.Base for this was the "desat alarm" function, a specific alarm behavior for monitoring of neonates which is enabled by default setting.This function increases the priority for a desaturation alarm from medium to high once the device detects a decrease of spo2 by 10% below the threshold.This function will however be disabled by activation of the satseconds alarm suppression feature; the ifu clearly refer in detail to that.Dräger finally concludes that there is no issue with the device which would require repair or correction.The observed behavior was related to the activation of a specific mode which disables another one.It is under responsibility and control of the user to adjust monitoring parameter to the specific needs of the individual patient and to verify appropriateness of changes after their execution.The dräger engineer explained his expertise to the staff and disabled the satseconds function after having agreed this with the users.Onsite support returned the iacs to satseconds off (default) which resolved the reported issue.This was a configuration/setting issue with no malfunction of the iacs and therefore this case has been closed.
|
|
Search Alerts/Recalls
|