MAUDE Adverse Event Report: APOLLO ENDOSURGERY, INC. ORBERA 365 INTRAGASTRIC BALLOON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY, INC. ORBERA 365 INTRAGASTRIC BALLOON Back to Search Results
Model Number B-50012
Device Problems Migration or Expulsion of Device (1395); Expulsion (2933)
Patient Problem Failure of Implant (1924)
Event Date 02/06/2023
Event Type  malfunction  
Manufacturer Narrative
Combined medwatch submitted to the fda on 21/jul/2023.A review of the device labeling notes the following: the current orbera365¿ intragastric balloon system directions for use (dfu) addresses the known and anticipated potential events of deflation (migration) and vomiting as follows: "the physiological response of the patient to the presence of the orbera365¿system balloon may vary depending upon the patient's general condition and the level and type of activity.The types and frequency of administration of drugs or diet supplements and the overall diet of the patient may also affect the response." "each patient must be monitored closely during the entire term of treatment in order to detect the development of possible complications.Each patient should be instructed regarding symptoms of deflation, gastrointestinal obstruction, ulceration and other complications which might occur, and should be advised to contact his/her physician immediately upon the onset of such symptoms." "complications: possible complications of the use of the orbera365¿ system include: gastric discomfort, feelings of nausea and vomiting following balloon placement as the digestive system adjusts to the presence of the balloon.Patients reporting loss of satiety, increased hunger and/or weight gain should be examined endoscopically as this is indicative of an igb deflation.A patient who's deflated (i.E.Collapsed) igb has moved into the intestines must be monitored closely for an appropriate period of time (at least 2 weeks) to confirm its uneventful passage through the intestine.Abdominal or back pain, either steady or cyclic.Deflation and subsequent replacement." deflated device should be removed promptly.Possible adverse events: intestinal obstruction by the igb.An insufficiently filled igb or a leaking igb that has lost sufficient volume may be able to pass from the stomach into the small bowel.It may pass all the way into the colon and be passed with stool.However, if there is a narrow area in the bowel or adhesion formation, which may occur after previous surgery on the bowel, the igb may not pass and could cause a bowel obstruction.If this occurs, surgery or endoscopic removal could be required.Insufficient or no weight loss.Igb deflation (i.E.Collapse) and subsequent replacement.Additional information: a device history record (dhr) review was performed due to the complaint being mdr reportable and the lot number was provided however the device was not returned.The subject product met all specifications and requirements in effect at the time of manufacture.There were no other complaints in the apollo database against this lot number af04891 and allegation.Device evaluation summary: assessment of the device involved in this complaint was not possible, and it has not been possible to determine the root cause for this event.
 
Event Description
Healthcare provider reported the patient vomited the balloon.Healthcare provider verified via abdomen echo the balloon was not found in the stomach or elsewhere.Patient is doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORBERA 365 INTRAGASTRIC BALLOON
Type of Device
INTRAGASTRIC BALLOON
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s. capital of texas hwy
bldg. 1 ste. 300
austin TX 78746
Manufacturer (Section G)
APOLLO ENDOSURGERY COSTA RICA, SRL
coyol free zone
bldg b 13.3
alajuela, cs CRI
CS   CRI
Manufacturer Contact
adriana russell
1120 s. capital of texas hwy
bldg. 1 ste. 300
austin, TX 78746
5122795114
MDR Report Key17370909
MDR Text Key320055885
Report Number3006722112-2023-00141
Device Sequence Number1
Product Code LTI
UDI-Device Identifier10811955020725
UDI-Public(01)10811955020725(17)20240203(10)AF04891
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P140008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/03/2024
Device Model NumberB-50012
Device Catalogue NumberB-50012
Device Lot NumberAF04891
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/27/2023
Initial Date FDA Received07/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age22 YR
Patient SexFemale
Patient Weight86 KG
-
-