• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH EXPEDIUM VERSE SPINE SYSTEM POLYAXIAL SCREW 5.5 7.0 X 35MM; ORTHOSIS, SPINAL PEDICLE FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH EXPEDIUM VERSE SPINE SYSTEM POLYAXIAL SCREW 5.5 7.0 X 35MM; ORTHOSIS, SPINAL PEDICLE FIXATION Back to Search Results
Catalog Number 199721735S
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/03/2023
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d2b: additional device product codes: osh, kwp, nkb, kwq, mnh.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j employee.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in the united kingdom as follows: it was reported that during surgery on (b)(6) 2023, the screw tab broke off while the screw was being inserted.The screw was removed and replaced with a different one without issue.There was no adverse outcome.This report involves one expedium verse spine system polyaxial screw 5.5 7.0 x 35mm.This is report 1 of 1 for (b)(4).
 
Event Description
The surgery was completed successfully with no delay.No fragments were left in the patient's body.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h4, h6 investigation summary : the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device from attachment.Visual analysis of the photo revealed that the 5.5 exp verse scr 7.0x35 was observed broken off from one tab.No other issues were found.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for 5.5 exp verse scr 7.0x35.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history: a review of the receiving inspection (ri) for 5.5 exp verse scr 7.0x35 was conducted identifying that lot number 246114 was released in one batch.Batch1: lot qty of 39 units were released on 25.July.2019 with no discrepancies.Supplier : (b)(4).As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPEDIUM VERSE SPINE SYSTEM POLYAXIAL SCREW 5.5 7.0 X 35MM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key17371713
MDR Text Key319591836
Report Number1526439-2023-01314
Device Sequence Number1
Product Code MNI
UDI-Device Identifier10705034444185
UDI-Public(01)10705034444185
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K142185
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number199721735S
Device Lot Number246114
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/21/2023
Initial Date FDA Received07/21/2023
Supplement Dates Manufacturer Received07/21/2023
08/17/2023
Supplement Dates FDA Received07/21/2023
08/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/25/2019
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-