Siemens requested more information and instrument data files for further investigation.Insufficient information was received from the customer.The cause of the hematocrit (hct) and hemoglobin (hgb) inconsistency could not be determined from the information provided by the customer.Epoc calculated hgb is derived from the hematocrit (hct) reading on the epoc device.As per section 12.16.6 of the epoc system manual; hemoglobin concentration is calculated from the measured hematocrit according to the relation: chgb (g/dl) = hct (decimal fraction) x 34.Given this relation, the investigation focused solely on hct performance of the card lot 03-23067-50.The in-house performance of hct for the card lot in question, 03-23067-50, did not identify any product deficiencies.Card lot 03-23067-50 was tested with blood and with aqueous control fluids at the time of product release.Card lot 03-23067-50 met product specifications at the time of release.Additional testing data was available on card lot 03-23067-50 and no issues were identified.Therefore, there is no evidence from in-house test data that the system or reagent cards are not performing as intended.The cause of this event is unknown.
|