BOSTON SCIENTIFIC CORPORATION OBTRYX SYSTEM - HALO; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, RETR Back to Search Results
Model Number M0068505000
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Neuropathy (1983); Pain (1994)
Event Date 06/22/2023
Event Type  Injury  
Manufacturer Narrative
Block h6: imdrf patient codes e2330 capture the reportable event of pain.Imdrf patient codes e0126 capture the reportable event of neuropathy.Imrdf impact code f2303 captures the reportable event of treatment with analgesics.Imrdf impact code f1905 captures the reportable event of device revision or replacement.
 
Event Description
It was reported to boston scientific corporation that an obtryx system halo device was implanted during a suburethral sling procedure performed on (b)(6) 2021.It was reported that after the placement of the sling, the patient experienced neuropathic-type pain at the anal, perineal, inner thigh, and outer hip levels.The patient was treated with analgesics.On (b)(6) 2023, a 7-centimeter portion of the sling was removed given the ineffectiveness of analgesic drug treatment.
 
Event Description
It was reported to boston scientific corporation that an obtryx system halo device was implanted during a suburethral sling procedure performed on (b)(6) 2021.It was reported that after the placement of the sling, the patient experienced neuropathic-type pain at the anal, perineal, inner thigh, and outer hip levels.The patient was treated with analgesics.On (b)(6) 2023, a 7-centimeter portion of the sling was removed given the ineffectiveness of analgesic drug treatment.
 
Manufacturer Narrative
Blocks d7a and h10 (block e1 note below) have updated based on the additional information received on august 2, 2023.Block e1: this event was reported by ch pitie salpetriere aphp where the partial removal of the mesh was performed.Block h6: imdrf patient codes e2330 capture the reportable event of pain.Imdrf patient codes e0126 capture the reportable event of neuropathy.Imrdf impact code f2303 captures the reportable event of treatment with analgesics.Imrdf impact code f1905 captures the reportable event of device revision or replacement.
 
Event Description
It was reported to boston scientific corporation that an obtryx system halo device was implanted during a suburethral sling procedure performed on (b)(6) 2021.It was reported that after the placement of the sling, the patient experienced neuropathic-type pain at the anal, perineal, inner thigh, and outer hip levels.The patient was treated with analgesics.On (b)(6) 2023, a 7-centimeter portion of the sling was removed given the ineffectiveness of analgesic drug treatment.
 
Manufacturer Narrative
Block h11: correction to block b5 - obtryx implantation date was corrected.Block e1: this event was reported by ch pitie salpetriere aphp where the partial removal of the mesh was performed.Block h6: imdrf patient code e2330 captures the reportable event of pain.Imdrf patient code e0126 captures the reportable event of neuropathy.Imrdf impact code f2303 captures the reportable event of treatment with analgesics.Imrdf impact code f1905 captures the reportable event of mesh revision.Block h10: upon receipt at our quality assurance laboratory, this obtryx system - halo underwent a thorough analysis.The device was visually inspected, and it revealed signs of normal usage, stretching and a part was missing.Since it was reported that the mesh was removed from the patient, the damage likely occurred during removal and should not be considered a failure.The product labeling review identified that the device was used per the directions for use (dfu) and that the adverse event of pain is a potential outcome of this procedure.The reported patient observations of neuropathy, medications and device revision are secondary effects of the other reported events or are unrelated to the device i.E., associated with patient factors, concomitant medical issues, concomitant medications, etc.Based on the information available and analysis results, the reported patient symptoms are known risks associated with obtryx system procedures and are noted as such in the device instructions for use.A conclusion code of known inherent risk of device was assigned to this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OBTRYX SYSTEM - HALO
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, RETR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
5086834015
MDR Report Key17374091
MDR Text Key319602983
Report Number3005099803-2023-03927
Device Sequence Number1
Product Code OTN
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K040787
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/01/2023
Device Model NumberM0068505000
Device Catalogue Number850-500
Device Lot Number0026111361
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/07/2023
Initial Date FDA Received07/21/2023
Supplement Dates Manufacturer Received08/02/2023
02/26/2024
Supplement Dates FDA Received08/24/2023
03/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-