The reagent pack used for the questionable results was first registered on the analyzer on 26-jun-2023 and no new calibration was performed.The customer had performed a lot calibration with a different reagent pack on (b)(6) 2023.The customer did not follow the reccomendations for calibration frequency.Product labeling states: "renewed calibration is recommended as follows: after 1 month (28 days) when using the same reagent lot.After 7 days (when using the same reagent kit on the analyzer).As required: e.G.Quality control findings outside the defined limits." the customer did not perform qc on the day of the event.Product labeling states: "controls for the various concentration ranges should be run individually at least once every 24 hours when the test is in use, once per reagent kit, and following each calibration." based on the information provided, the specific cause of the event could not be determined.A general reagent kit or lot issue is not suspected, as the qc with the relevant kit was acceptable on the date the kit was registered, and another kit from the same reagent lot performed well.The investigation determined the event was consistent with a use error at the customer site.
|