MAUDE Adverse Event Report: KARL STORZ SE & CO. KG ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR Back to Search Results
Model Number 011050-10
Device Problems Break (1069); Fracture (1260)
Patient Problem Insufficient Information (4580)
Event Date 07/04/2023
Event Type  Injury  
Event Description
It was reported that the curved part of the handle has broken and detached in the patient's uterus.The part could not be found.It is considered that it was removed during the abrasion of the uterus during the removal of the compresses.Another device (of a different lot) was used to complete the intervention.This was therefore extended by 30 minutes.Since it is not confirmed that the broken part was removed, a report is required.
 
Manufacturer Narrative
The item in question was returned to the manufacturer.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
 
Manufacturer Narrative
The investigation report was completed on august 2, 2023 due to the ductile break of the cutting loop, it is most likely that the loop got mechanically overloaded, broke, and created a shortcut to the neutral electrode resulted in melting down the wire.There is no material missing, just molten down and deposit on the neutral electrode.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR
Type of Device
ELECTRODE, BIPOLAR
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245
MDR Report Key17379327
MDR Text Key319656885
Report Number9610617-2023-00175
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K882270
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number011050-10
Device Catalogue Number011050-10
Device Lot Number839167
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/10/2023
Initial Date FDA Received07/24/2023
Supplement Dates Manufacturer Received08/02/2023
Supplement Dates FDA Received08/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/19/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-