MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VALITUDE CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VALITUDE CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number U125
Device Problems Pacing Problem (1439); Inaccurate Synchronization (1609); Under-Sensing (1661)
Patient Problem Arrhythmia (1721)
Event Date 07/08/2023
Event Type  Injury  
Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) exhibited ventricular undersensing.This resulted in additional pacing that was suspected to have triggered non-sustained ventricular tachycardia (nsvt).It was noted that the crt-p was programmed to a higher fixed rv sensitivity.Technical services (ts) reviewed troubleshooting options, including adjusting rv sensitivity.This crt-p remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALITUDE CRT-P
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17386463
MDR Text Key319654334
Report Number2124215-2023-39192
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559389
UDI-Public00802526559389
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/28/2022
Device Model NumberU125
Device Catalogue NumberU125
Device Lot Number717396
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/17/2023
Initial Date FDA Received07/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/19/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age88 YR
Patient SexFemale
-
-