• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problems Microbial Contamination of Device (2303); Insufficient Information (3190)
Patient Problems Bacterial Infection (1735); Insufficient Information (4580)
Event Date 11/16/2016
Event Type  Death  
Event Description
The event was identified through a retrospective review of product liability lawsuits.Through this review, we have identified a few events where the plaintiff had filed a suit against the company, and it was handled by our legal department, but the underlying product information had not been forwarded to the complaint handling unit.Livanova opened a capa to identify any possible lawsuits that might require complaint reporting and to retrospectively submit those events that were not previously submitted and to prevent future similar issues from occurring.Complainant alleges through their counsel that he/she underwent an operation on (b)(6), 2016.No additional information was provided.Based on current status of the investigation the alleged device issue was yet not confirmed.
 
Manufacturer Narrative
A.1.-a.5.Patient information was not provided.D.4.Serial number is unknown.This information will be provided in a supplemental report if made available.H.4.As the serial number is unknown, the device manufacture date could not be determined.This information will be provided in a supplemental report if made available.G.5.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H.9.Livanova deutschland implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in bristol, united kingdom.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial report.
 
Manufacturer Narrative
H10: through follow up communication, livanova learned that alleged malfunction claimed by the complainant was a mycobacteria chimaera infection acquired following a heart valve surgery in which a heater-cooler system 3t device was used.In addition, it was learned that possible involved device serial numbers could be the following: (b)(6) (model: 16-02-80).No other information has been made available from the customer.Device history records (dhrs) review of possible involved device serial numbers has been completed and did not identify any deviation or non-conformity relevant to the reported issue.No device, between the ones possible involved and in use at the hospital at the time of surgery (2016), was upgraded with vacuum and sealing kit.No further investigation is possible.Source of patient contamination remains unknown and a direct relationship between the reported adverse event and the device could not be established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key17388803
MDR Text Key319675482
Report Number9611109-2023-00351
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 03/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number16-02-80
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/17/2020
Initial Date FDA Received07/25/2023
Supplement Dates Manufacturer Received02/05/2024
Supplement Dates FDA Received03/04/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2076/2081-2015
Patient Sequence Number1
Patient Outcome(s) Death;
-
-