MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPT; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number B35200 |
Device Problems
Break (1069); High impedance (1291); Difficult to Insert (1316)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/21/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Section d references the main component of the system.Other medical products in use during the event include: brand name activa; product id: 3389; product type: 0200-lead; implant date: (b)(6) 2023; brand name: activa; product id: 3708660; product type: 0191-extension; implant date (b)(6) 2021; explant date: (b)(6) 2023, brand name: activa; product id: 3708660; product type: 0191-extension; implant date: (b)(6) 2021; explant date: (b)(6) 2023.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the healthcare provider (hcp) and patient (pt) wanted to be mri eligible.The extensions were exchanged.The hcp attempted to test with twist lock but was unable to insert stylet.Used alligator clips which detected high impedance on electrodes 0, 1, and 3 which was also displayed when the implantable neurostimulator (ins) was interrogated.The lead also had an exposed wire and future lead replacement is planned.Patient was still receiving therapy with no side effects displayed.No symptoms reported.
|
|
Event Description
|
Additional information receivedfrom the manufacturer¿s representative (rep) reported the patient was receiving adequate therapy from remaining contact two (which was the only contact without high impedances), so the lead was going to remain in use.The cause of the high impedance and stylet issue was due to the stylet being unable to be inserted to utilize the twist lock cable.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|