• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ENSITE VELOCITY¿ DWS7 DISPLAY WORKSTATION; ELECTROPHYSIOLOGY CARDIAC MAPPING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ENSITE VELOCITY¿ DWS7 DISPLAY WORKSTATION; ELECTROPHYSIOLOGY CARDIAC MAPPING SYSTEM Back to Search Results
Model Number 100129557
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/28/2023
Event Type  malfunction  
Event Description
During a paroxysmal atrial fibrillation, the ensite velocity display workstation computer would not turn on which resulted in a delay of 3 hours.The connections were checked and the ensite velocity display workstation computer was restarted which did not resolve the issue.The ensite velocity display workstation computer was replaced, and the procedure was completed with no adverse consequences to the patient.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported power issue could not be determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE VELOCITY¿ DWS7 DISPLAY WORKSTATION
Type of Device
ELECTROPHYSIOLOGY CARDIAC MAPPING SYSTEM
Manufacturer (Section D)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key17389812
MDR Text Key319890020
Report Number2184149-2023-00147
Device Sequence Number1
Product Code DQK
UDI-Device Identifier05415067014573
UDI-Public05415067014573
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K172396
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number100129557
Device Lot Number7355182
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/28/2023
Initial Date FDA Received07/25/2023
Supplement Dates Manufacturer Received10/18/2023
Supplement Dates FDA Received10/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/27/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-