Model Number EARLYVUE VS30 VITAL SIGNS MONITOR |
Device Problems
Display or Visual Feedback Problem (1184); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/29/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).A follow up report will be submitted once the investigation is complete.
|
|
Event Description
|
It was reported that the system hangs during measurement.The device was in use at time of event, there was no adverse event reported.
|
|
Manufacturer Narrative
|
The device was sent to philips bench repair.The bench repair technician (brt) found that the device would hang occasionally and determined that the main board required replacement.The brt replaced the main board and performed the stk and function check, and the device passed.The device was operational after replacing the main board was completed.
|
|