This report has been identified as b.Braun medical internal report number (b)(4).One (1) sample was submitted to the manufacturer for evaluation.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.Through visual examination, it was observed that the safety clip was not engaged, and the cannula was bent.The safety clip had dislodged and the backwall was curved.The clip hole was inspected under 20x magnification, and no damages were observed.The cannula surface was inspected, and clip travel scratched lines were observed along the cannula.Cannula od was measured and the od measurements were consistent along the cannula and they are within specification.A simulation was performed on a reserved sample to replicate the defect.In the simulation, the cannula was withdrawn in a slanted angle (not following instruction for use), then the safety clip was dislodged, the clip arm deformed and the backwall was curved outward as a "v-shape" upon reaching the tip.The bending angle of the cannula was almost similar to the complaint sample.Based on the investigation, the defect is likely due to an application error.Therefore, we consider the complaint not confirmed.
|