• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Catalog Number B5ST
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/05/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 7/27/2023.D4: batch # unk.A manufacturing record evaluation is pending.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that during a laparoscopic ob-gyn procedure, the inner cylinder was bent when the trocar was punctured.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Manufacturer Narrative
(b)(4), date sent: 8/28/2023.D4: batch # a9c715.The product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that only the obturator from the b5st device was returned with the obturator cannula damaged bent.No conclusion could be reached as to what might have caused the damage observed at analysis.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH XCEL BLADELESS TROCAR
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key17405887
MDR Text Key320251954
Report Number3005075853-2023-05352
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10705036001232
UDI-Public10705036001232
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberB5ST
Device Lot Number310C90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2023
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/06/2023
Initial Date FDA Received07/27/2023
Supplement Dates Manufacturer Received08/02/2023
Supplement Dates FDA Received08/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/02/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-