• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ROTAPRO; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ROTAPRO; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 39467-150
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/05/2023
Event Type  malfunction  
Event Description
It was reported that a saline leak occurred.The 95% stenosed target lesion was in the moderately tortuous and severely calcified distal right coronary artery.A 1.50mm rotapro was selected for use.The device was first platformed with no issue outside the body before advancing into the patient.During the procedure, the physician had difficulty passing the lesion with the burr and after four minutes, it was noted that the burr started slowing down to 130,000rpm doing the same motion.After the burr successfully was through the lesion, it was taken out successfully.However, while using dynaglide mode, the physician noticed that the shaft of the burr had blood in it and saline was leaking in the middle portion of the burr catheter.Hence, it was determined that the leak was caused by the pushing of the whole burr along the advancer knob.The procedure was completed with the same 1.5mm burr.There was no patient complication reported.
 
Manufacturer Narrative
E1.Initial reporter address 1-800 w central texas expy suite 355.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAPRO
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
model farm road
cork
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17406818
MDR Text Key320158838
Report Number2124215-2023-37306
Device Sequence Number1
Product Code MCX
UDI-Device Identifier08714729893356
UDI-Public8714729893356
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number39467-150
Device Catalogue Number39467-150
Device Lot Number0031218862
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/05/2023
Initial Date FDA Received07/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/13/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-