• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL BRK¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL BRK¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH Back to Search Results
Model Number G407208
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/06/2023
Event Type  malfunction  
Event Description
During an atrial fibrillation procedure, during transseptal puncture it was noted that the distal part of the needle was broken.After multiple attempts to insert the needle the distal part of the needle was noted to be on the prep table.There were no adverse patient health consequences.
 
Manufacturer Narrative
One brk transseptal needle was received for evaluation.The brk needle had been fractured at its distal end.The brk needle instructions for use state ¿do not alter this device in any way¿.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the needle fracture is consistent with not following the instructions for use.
 
Manufacturer Narrative
After further review, analysis confirmed the lot number of the device to be 9050029.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH
Type of Device
TRANSSEPTAL NEEDLE
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key17411065
MDR Text Key319980682
Report Number3008452825-2023-00321
Device Sequence Number1
Product Code DRC
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K122587
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG407208
Device Lot Number9050029
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/07/2023
Initial Date FDA Received07/27/2023
Supplement Dates Manufacturer Received09/05/2023
09/05/2023
Supplement Dates FDA Received09/13/2023
09/13/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-