SCHILLER AG TEMPUS LS-MANUAL; LOW ENERGY DEFIBRILLATOR Back to Search Results
Model Number 00-3020
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/10/2022
Event Type  malfunction  
Event Description
This report is based on information provided by philips remote service engineer and schiller investigation team and has been investigated by the philips complaint handling team.Philips received a complaint on the tempus ls, indicating that device dpm hardware failure occurred during pacing.The complaint was escalated for technical investigation and the results indicate that the pacer board was not glued ¿protection folio of the vhb facing the igbt module has not been removed before dpm assembly.It has been concluded that the failure is due to an production operator error.In order to avoid this kind of failure a verification process must be included once both boards have been assembled.The manufacturer is implementing manual verification of the mechanical maintaining of the pacer board (wsm0218a) after assembly of both boards (wsm0217a and wsm0218a).This will be added in an updated dpm assembly document : dmind 8188.The device moved to loaner pool after the dpm glued correctly.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.The investigation concludes that no further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS LS-MANUAL
Type of Device
LOW ENERGY DEFIBRILLATOR
Manufacturer (Section D)
SCHILLER AG
altgasse 68
baar 6341
SZ  6341
Manufacturer (Section G)
SCHILLER AG
altgasse 68
baar 6341
SZ   6341
Manufacturer Contact
jon yard
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 -6XW
UK   GU14 6XW
MDR Report Key17413225
MDR Text Key319986572
Report Number3003832357-2023-00495
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number00-3020
Device Catalogue Number989706001681
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 08/10/2022
Initial Date FDA Received07/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/12/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-