• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: REMOTE DIAGNOSTIC TECHNOLOGIES LTD. TEMPUS PRO; MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

REMOTE DIAGNOSTIC TECHNOLOGIES LTD. TEMPUS PRO; MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS) Back to Search Results
Model Number 00-1024-R
Device Problem Power Problem (3010)
Patient Problem Chest Pain (1776)
Event Date 07/11/2022
Event Type  Injury  
Event Description
It has been reported to philips that the monitor did not start up with a patient with chest pain.This report is based on information provided by philips repair service personnel and has been investigated by the philips complaint handling team.Philips received a complaint on the tempus pro indicating that the monitor did not start up with a patient with chest pain.The complaint was escalated for technical investigation and the results indicate that there is a problem with the system board and software needs an upgrade.Upgrade 4.30, functional check and safety test.Rdt rainbow system board assembly w/nibp got replaced and the software was upgraded to 4.30.Functional check and safety test performed successfully.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.The device was operational after repairs were completed.The investigation concludes that no further action is required at this time.
 
Manufacturer Narrative
Updated type of complaint from product problem to serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS PRO
Type of Device
MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS)
Manufacturer (Section D)
REMOTE DIAGNOSTIC TECHNOLOGIES LTD.
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 6XW
UK  GU14 6XW
Manufacturer (Section G)
REMOTE DIAGNOSTIC TECHNOLOGIES LTD.
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 6XW
UK   GU14 6XW
Manufacturer Contact
tanya deschmidt
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 -6XW
UK   GU14 6XW
MDR Report Key17413323
MDR Text Key320241998
Report Number3003832357-2023-00496
Device Sequence Number1
Product Code MHX
UDI-Device Identifier05060472441027
UDI-Public05060472441027
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K201746
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number00-1024-R
Device Catalogue Number989706000051
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/11/2022
Initial Date FDA Received07/28/2023
Supplement Dates Manufacturer Received07/11/2022
Supplement Dates FDA Received11/22/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-