• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ROTAWIRE DRIVE; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ROTAWIRE DRIVE; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 2077-01
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/07/2023
Event Type  malfunction  
Event Description
It was reported that a contamination occurred.A rotawire drive was selected for use.When the device was taken out of the hoop, it was noted that a rough substance was attached at the bottom part of the wire.The device was not used inserted inside the patient.The procedure was completed with another of the same device.There were no patient complications reported.
 
Event Description
It was reported that a contamination occurred.A rotawire drive was selected for use.When the device was taken out of the hoop, it was noted that a rough substance was attached at the bottom part of the wire.The device was not used inserted inside the patient.The procedure was completed with another of the same device.There were no patient complications reported.
 
Manufacturer Narrative
Device evaluated by mfr: the device was returned for evaluation.Microscope inspection revealed that the spring tip was detached.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAWIRE DRIVE
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17414926
MDR Text Key319987514
Report Number2124215-2023-37584
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P900056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2077-01
Device Catalogue Number2077-01
Device Lot Number0030369660
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/07/2023
Initial Date FDA Received07/28/2023
Supplement Dates Manufacturer Received08/31/2023
Supplement Dates FDA Received09/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/18/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-