Model Number 980X1ENDIUU
Device Problems Defective Alarm (1014); No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2023
Event Type  malfunction  
Event Description
It was reported that while in use on a patient, the 980 ventilator shutdown with no alarms. patient was removed from the ventilator and placed on an alternate ventilator with no injury.
 
Manufacturer Narrative
H3 device evaluation summary: medtronic conducted an investigation based upon all information received.It was reported that while in use on a patient, the 980 ve ntilator shutdown with no alarms.  the device was available for evaluation.Review of logs confirmed the unit had an intermittent shutdown leading to loss of ventilation.The customer found batteries completely depleted and stated that unit must have lost alternating current (ac) power.The customer charged the batteries and found them to function properly.Additionally it was also found batteries nearing end of life so customer replaced batteries themselves.The medtronic service personnel (sp) checked the unit and was unable to duplicate the reported issue.The unit passed all test and calibrations as per manufacturer specifications at the time of service.The cause of the observed condition could not be determined.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it has met all medtronic quality s pecifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
980 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
EI 
Manufacturer (Section G)
COVIDIEN
micheal collins rd mervue
galway
EI  
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key17415979
MDR Text Key320215335
Report Number8020893-2023-00433
Device Sequence Number1
Product Code CBK
UDI-Device Identifier10884521171541
UDI-Public10884521171541
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131252
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number980X1ENDIUU
Device Catalogue Number980X1ENDIUU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/03/2023
Initial Date FDA Received07/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/16/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-