SCHILLER AG TEMPUS LS-MANUAL; LOW ENERGY DEFIBRILLATOR Back to Search Results
Model Number 00-3020
Device Problem No Pacing (3268)
Patient Problem Chest Pain (1776)
Event Date 05/09/2022
Event Type  Injury  
Event Description
Philips received a complaint on the tempus ls indicating that the customer tried to pace a patient and the device communicated hardware failure.The schiller investigated the reported problem based on the provided logfiles of the device and confirmed log file showed that the issue is most likely to be related to an intermittent "error 26" issue.The pacer stopped due to the error 26.However, the trigger and actual root cause of the error 26 cannot be concluded at this stage.The defective device sent to the loaner pool for dpm rework at schiller.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.The investigation concludes that no further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS LS-MANUAL
Type of Device
LOW ENERGY DEFIBRILLATOR
Manufacturer (Section D)
SCHILLER AG
altgasse 68
baar 6341
SZ  6341
Manufacturer (Section G)
SCHILLER AG
altgasse 68
baar 6341
SZ   6341
Manufacturer Contact
tanya deschmidt
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 -6XW
UK   GU14 6XW
MDR Report Key17429661
MDR Text Key320114650
Report Number3003832357-2023-00505
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number00-3020
Device Catalogue Number989706001681
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 05/09/2022
Initial Date FDA Received07/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-