• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC 7MM EXTENDED LENGTH ENDOSCOPE; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC 7MM EXTENDED LENGTH ENDOSCOPE; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number VH-1111
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/30/2023
Event Type  malfunction  
Manufacturer Narrative
Trackwise id (b)(4).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event Description
The hospital reported that during repossessing, 7mm extended length endoscope s/n (b)(6) had a crack in the lens, cleaning it does not work.This issue was found by the sterilization team during the sterilization process and was brought to their attention.No patient effects reported, as there was no patient involvement in this situation.
 
Manufacturer Narrative
Trackwise#: (b)(4).The device was returned to the factory for evaluation on 07/17/2023.An investigation was conducted on 08/01/2023.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.There were no visual defects observed on the endoscope.An image quality inspection was performed with an endoscopic video imaging system.A clear image was able to be obtained.The image was observed to be clear with no cracks observed.Based on the results of the evaluation, the reported failure "crack " was not confirmed.The reported device is an oem device.The certificate of conformance was reviewed for the serial#: (b)(6).The vendor certifies that this device lot conforms to all applicable product specifications and there were no non-conformances identified for the manufacturing batch.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
7MM EXTENDED LENGTH ENDOSCOPE
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key17431289
MDR Text Key320170382
Report Number2242352-2023-00625
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier00607567700802
UDI-Public00607567700802
Combination Product (y/n)N
PMA/PMN Number
K014250
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVH-1111
Device Catalogue NumberVH-1111
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/05/2023
Initial Date FDA Received07/31/2023
Supplement Dates Manufacturer Received09/15/2023
Supplement Dates FDA Received09/20/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
N/A.
-
-