MEDTRONIC IRELAND ENDURANT IIS BIFURCATED STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
|
Back to Search Results |
|
Model Number ESBF3614C103EJ |
Device Problem
Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/20/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
An endurant iis bifurcate stent graft was intended to be implanted in the endovascular treatment of a 56mm aaa. it was reported that during the index procedure, when checking the contralateral side gate with the "e" marker before inserting inside the main device, the "e' mark appeared distorted; the "e" marker did not match the contralateral side gate mark and blue dot, so the contralateral side gate's deployment direction could not be confirmed. not only did cannulation become more difficult when opened on the contralateral side gate, but the shape of the aneurysm also made the stent graft patency after placement uneasy, meaning the lumen of the aneurysm was crushed and could be pushed depending on the position where the contralateral gate is deployed. it was determined to be a defective device, and a different product was used. the procedure was completed without any problems using a different product.It was confirmed the ifu was followed.Per the physician the cause was a defective product. no additional clinical sequalae were provided and the patient is fine.
|
|
Manufacturer Narrative
|
Product analysis device returned with the external slider and backend wheel in the home position.There was no deformation evident to the device.The ¿e¿ marker was not visible due to the folding of the material when the graft was loaded.Fluoroscopy imaging confirmed the ¿e¿ marker was not distorted.On bench top deployment of the graft no abnormality was evident to the ¿e¿ marker, flow divider marker and gate markers positioning on the graft.The reported dimensional deviation could not be confirmed through analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|