Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/06/2023 |
Event Type
malfunction
|
Event Description
|
It was reported the 3.5 hex driver broke as the surgeon was loosening the femoral screw.No pieces fell into the patient and there was no patient impact.No addtional information is available.
|
|
Manufacturer Narrative
|
(b)(4).The product has been received by zimmerbiomet.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified sign of use and tip was fractured/broken.The device history record was reviewed and no discrepancies relevant to the reported event were found.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Medical records were not provided.The root cause of the reported issue is attributed to wear and tear resulting from use.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|