• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. FINN SEGMENT HEX DRIVER; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. FINN SEGMENT HEX DRIVER; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/06/2023
Event Type  malfunction  
Event Description
It was reported the 3.5 hex driver broke as the surgeon was loosening the femoral screw.No pieces fell into the patient and there was no patient impact.No addtional information is available.
 
Manufacturer Narrative
(b)(4).The product has been received by zimmerbiomet.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified sign of use and tip was fractured/broken.The device history record was reviewed and no discrepancies relevant to the reported event were found.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Medical records were not provided.The root cause of the reported issue is attributed to wear and tear resulting from use.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FINN SEGMENT HEX DRIVER
Type of Device
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17436227
MDR Text Key320159751
Report Number0001825034-2023-01785
Device Sequence Number1
Product Code HWR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number32-471247
Device Lot Number315470
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/10/2023
Initial Date FDA Received08/01/2023
Supplement Dates Manufacturer Received08/23/2023
Supplement Dates FDA Received08/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-