Model Number CLV-S190 |
Device Problem
Device Alarm System (1012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/03/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation, and the customer¿s allegation was not confirmed.The evaluation found the following: hinge assembly is not retaining its position properly, causing the high function to not work, there was rust on the hinge assembly, dust was found inside the device, tracks of the flat cable were rusted, heavy corrosion found on the rear panel and its components, and the power supply unit was faulty causing the unit to not turn on properly.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Event Description
|
The customer reported to olympus, the visera elite xenon light source had its front panel blinking during the daily inspection by the biomedical engineer on staff.The issue was found during maintenance with no procedure involved.There were no reports of patient harm.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the front panel blinking could not be identified, nor could the phenomenon be confirmed/reproduced.Also, based on the results of the legal manufacturer's investigation, a definitive root cause of the faulty power supply unit could not be identified.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|