Catalog Number 5F061503C |
Device Problems
Material Deformation (2976); Positioning Problem (3009)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/07/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the lifestent 5f vascular stent that are cleared in the us.The pro code and 510 k number for the lifestent 5f vascular stent are identified in d2 and g4.H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiry date: 07/2024).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during an angioplasty procedure via a contralateral approach, the shaft of the catheter allegedly kinked and was unable to reach the lesion.The procedure was completed by using another device.There was no reported patient injury.
|
|
Event Description
|
It was reported that during an angioplasty procedure via a contralateral approach, the shaft of the catheter allegedly kinked and was unable to reach the lesion.The procedure was completed by using another device.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the lifestent 5f vascular stent that are cleared in the us.The pro code and 510 k number for the lifestent 5f vascular stent are identified in d2 and g4.H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issues.Investigation summary: the sample was not returned for evaluation and no x-ray images were provided which leads to inconclusive evaluation result.A kink was reported but it was not known whether the kink caused the issue or whether the advancing difficulty caused the kink.Based on the investigation of the provided information, the investigation is closed as inconclusive.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.The instruction for use state: 'if resistance is met during stent system introduction, the stent system should be withdrawn and another stent system should be used.', and 'examine the stent system to ensure it has not been damaged during shipment and that its size, shape and condition are suitable for the procedure for which it is to be used.If it is suspected that the sterility or performance of the stent system has been compromised, the device should not be used.' the instruction for use further state: 'keep the device as straight as possible following removal from the packaging and while inserted in the patient.', and 'pre-dilatation of the lesion with a balloon dilatation catheter is recommended.' h10: d4 (expiry date: 07/2024), g3, h6 (device).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|