Model Number OER-ELITE |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
During the inspection of the returned asset device, it was found that the software version was 1.08 and would be upgraded to 1.09.No other malfunctions were found.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
|
Event Description
|
A user facility returned the olympus asset, endoscope reprocessor, to the olympus service center.Upon inspection and testing of the returned device, it was observed that the serial harness had a short in the connector for the central processing unit (cpu) j3.This report is being submitted for the event found during evaluation.There was no patient involvement.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Correction to e2, e3.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely the harness for the lid opening/closing sensor had a contact failure in central processing unit (cpu) board resulting in a failure to turn on the subject device.Olympus will continue to monitor field performance for this device.
|
|