• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NOVOCURE GMBH OPTUNE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NOVOCURE GMBH OPTUNE Back to Search Results
Model Number TFH9100
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Convulsion/Seizure (4406)
Event Date 07/06/2023
Event Type  Injury  
Manufacturer Narrative
Novocure medical opinion is that the seizure was related to the underlying disease and unrelated to optune therapy.Risk factors for seizure in this patient include: concomitant temozolomide (convulsions are listed as among the most common adverse reactions.Source: temozolomide prescribing information) and concomitant dexamethasone (convulsion is listed as a known adverse reaction.Source: dexamethasone prescribing information).Seizures were reported as adverse events on the pivotal phase iii recurrent gbm trial in both arms of the trial (9% and 4% in optune therapy and chemotherapy arms respectively).None of these seizures were considered device or chemotherapy related by investigators.Seizures are a known complication of gbm and have been reported as the presentation symptom in 27% of cases.During the course of the disease, 51% of patients will experience seizures (clin neurol neurosurg.2015; 139:166-171).
 
Event Description
A 40-year-old male with a high grade glioma started optune therapy on (b)(6) 2023.Novocure was informed on (b)(6) 2023, that the patient was hospitalized earlier that day due to a seizure.The patient noted he had discontinued an unspecified seizure medication the month prior.The patient's health care provider (hcp) was made aware and optune therapy was temporarily discontinued.On (b)(6) 2023, the hcp reported that the patient did not have a history of seizure.In addition, levetiracetam was previously prescribed during his hospital admission in (b)(6) 2023.In the emergency department (ed) following the seizure the patient was administered intravenous (iv) levetiracetam 1 gram and upon hospital admission 500mg iv every 12 hours.Lorazepam was prescribed as needed for breakthrough seizures.The patient was discharged on (b)(6) 2023, on oral levetiracetam 500mg bid, and advised to resume dexamethasone and follow up with radiation oncology and neurosurgery.The hcp assessed that the exact cause of the event was unknown, the event may or may not have involved underlying disease and/or optune therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTUNE
Type of Device
OPTUNE
Manufacturer (Section D)
NOVOCURE GMBH
business village d4
park 6/platz 10
root d4, 6039
SZ  6039
Manufacturer (Section G)
NOVOCURE GMBH
business village d4
park 6/platz 10
root d4, 6039
SZ   6039
Manufacturer Contact
sharon perez
195 commerce way
portsmouth, NH 03801
2077527602
MDR Report Key17445552
MDR Text Key320252962
Report Number3010457505-2023-00242
Device Sequence Number1
Product Code NZK
UDI-Device Identifier07290107982221
UDI-Public07290107982221
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100034
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 08/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberTFH9100
Device Catalogue NumberTFH9100US
Device Lot NumberN/A
Initial Date Manufacturer Received 07/06/2023
Initial Date FDA Received08/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/25/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
DEXAMETHASONE; FAMOTIDINE; LEVETIRACETAM; METHOCARBAMOL; OXYCODONE-ACETAMINOPHEN
Patient Outcome(s) Hospitalization;
Patient Age40 YR
Patient SexMale
Patient Weight95 KG
-
-