|
Catalog Number RED72SDKIT |
Device Problems
Fracture (1260); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/09/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation of the returned red72 confirmed a fracture near the hub, underneath the strain relief.If the device is mishandled during manipulation against resistance, damage such as a fracture may occur.Due to the fracture, the device could not be functionally tested, and the root cause of resistance during the procedure could not be determined.Further evaluation revealed a kink on the catheter shaft.This damage was likely incidental to the reported complaint and may have occurred during return to penumbra.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Event Description
|
The patient was undergoing a thrombectomy procedure using a penumbra system red 72 reperfusion catheter (red72), a sendit delivery device (sendit), and a non-penumbra sheath.It was reported that the patient¿s anatomy was tortuous.During the procedure, the red72 and the sendit were flushed and soaked in saline prior to use.The physician advanced the red72 and sendit to the target location.When the physician attempted to retract the sendit, the sendit would not move as it was stuck in the red72.The physician fractured the red72 at the strain relief below the proximal hub when attempting to remove the sendit.The red72 and the sendit were removed from the patient.The procedure was completed using another red72, a penumbra system red 43 reperfusion catheter (red43), and the same sheath.There was no report of an adverse effect to the patient.
|
|
Search Alerts/Recalls
|
|
|