The patient was undergoing a coil embolization procedure in the splenic vein using a lantern delivery microcatheter (lantern), ruby coils, a non-penumbra guide catheter and a guidewire.During the procedure, the physician implanted three ruby coils in the target location using the lantern.While attempting to remove the pusher assembly of the last coil from the lantern, the physician noticed that the proximal end of the lantern was fractured and the mid-shaft was unraveled.It was reported that the lantern was fractured but remained connected by a thread.Therefore, the physician removed the lantern by pulling it out.The procedure ended at this point and the vessel was completely embolized.There was no report of an adverse effect to the patient.
|
Evaluation of the returned lantern confirmed that the catheter was fractured.Evaluation revealed that the support coil winds were exposed at the fractured location.If the lantern is retracted at an angle, damage such as a kink and subsequent fracture may occur.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|