REMOTE DIAGNOSTIC TECHNOLOGIES LTD. TEMPUS PRO; MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS) Back to Search Results
Model Number 00-1024-R
Device Problem Intermittent Loss of Power (4016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported to philips while transferring a patient, monitor froze and restarted.Tried to do 12 lead and froze again and shut down.
 
Manufacturer Narrative
Diagnostics determined the device would require return for repair and it was returned to a philips bench repair facility.The reported complaint was confirmed as the unit would not re-boot correctly due to a failure with the 4g firmware and would not load software.The rear cover assembly (rdt rear cover with printer kit (for 00-, part 453564865051) was replaced resolving the customer¿s complaint.Testing and verification was completed satisfactorily (calibrated nbp, co2 and touch screen) and the device was returned to service at the customer site.Based on the information available and testing conducted, the cause of the reported problem was the 4g firmware.The reported problem was confirmed.Based on the information available, no further action is necessary at this time.The device was operational after repairs were completed.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS PRO
Type of Device
MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS)
Manufacturer (Section D)
REMOTE DIAGNOSTIC TECHNOLOGIES LTD.
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 6XW
UK  GU14 6XW
Manufacturer (Section G)
REMOTE DIAGNOSTIC TECHNOLOGIES LTD.
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 6XW
UK   GU14 6XW
Manufacturer Contact
tanya deschmidt
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 -6XW
UK   GU14 6XW
MDR Report Key17446273
MDR Text Key320387095
Report Number3003832357-2023-00507
Device Sequence Number1
Product Code MHX
UDI-Device Identifier05060472441027
UDI-Public05060472441027
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201746
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number00-1024-R
Device Catalogue Number989706000051
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/21/2023
Initial Date FDA Received08/02/2023
Supplement Dates Manufacturer Received08/05/2023
Supplement Dates FDA Received10/06/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-