Diagnostics determined the device would require return for repair and it was returned to a philips bench repair facility.The reported complaint was confirmed as the unit would not re-boot correctly due to a failure with the 4g firmware and would not load software.The rear cover assembly (rdt rear cover with printer kit (for 00-, part 453564865051) was replaced resolving the customer¿s complaint.Testing and verification was completed satisfactorily (calibrated nbp, co2 and touch screen) and the device was returned to service at the customer site.Based on the information available and testing conducted, the cause of the reported problem was the 4g firmware.The reported problem was confirmed.Based on the information available, no further action is necessary at this time.The device was operational after repairs were completed.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|