• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S GENESIS MALLEABLE PENILE PROSTHESIS; PENILE PROSTHESIS, ROD/CYLINDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S GENESIS MALLEABLE PENILE PROSTHESIS; PENILE PROSTHESIS, ROD/CYLINDER Back to Search Results
Catalog Number 91-9509SC / 519250
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2023
Event Type  Injury  
Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release and no trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Event Description
According to available information, this device required replacement due to a wire break.No other adverse patient effects were reported.
 
Manufacturer Narrative
Examination of the device revealed that rod 1 was unable to maintain the retention angle.The rod was bent in four directions into approximately 45-degree angle but did not hold angle.With use over time, the coil inside the device may fatigue which could result in the loss of angle retention.Based on the testing, it was confirmed that the inner core of rod 1 most likely fractured.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
 
Event Description
According to the available information this genesis malleable revised due malfunction (wire breakage).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GENESIS MALLEABLE PENILE PROSTHESIS
Type of Device
PENILE PROSTHESIS, ROD/CYLINDER
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key17446437
MDR Text Key320277984
Report Number2125050-2023-00913
Device Sequence Number1
Product Code FAE
UDI-Device Identifier05708932710831
UDI-Public5708932710831
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K040959
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number91-9509SC / 519250
Device Lot Number8493967_5192502022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/02/2023
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/03/2023
Date Device Manufactured04/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age61 YR
-
-