Catalog Number 4123201
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problems Dyspnea (1816); Hypersensitivity/Allergic reaction (1907); Paresthesia (4421); Cramp(s) /Muscle Spasm(s) (4521)
Event Date 06/21/2023
Event Type  Injury  
Manufacturer Narrative
Investigation is in process, a follow-up report will be provided.
 
Event Description
The customer reported severe citrate-related hypocalcaemia symptoms such as paresthesia, tetany and dyspnoea during a in stimulated leukapheresis procedure on spectra optia.The patient without known allergies, risk factors or pre-existing illness had this reaction from acd-a.Per follow up with the customer, the patient was given calcium and iv magnesium.The collection set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.10.Investigation: a disposable complaint history search was performed for this lot and found one additional report for similar issues on this lot, also located at robert-bosch-krankenhaus.See mdr 1722028-2023-00258.According to therapeutic apheresis: a physician's handbook, adverse events occur during therapeutic procedures with a frequency of 4.8%.Transient hypocalcemia associated with apheresis is usually well tolerated.Symptoms often show as parasethesia (tingling) but patients may also experience unusual taste, nausea, lightheadedness, shivering, and tremors.Severe hypocalcemia may also cause muscle contractions and can progress to tetany and seizures if hypocalcemia escalates and is not corrected.Investigation is in process, a follow-up report will be provided.
 
Event Description
The customer reported severe citrate-related hypocalcaemia symptoms such as paresthesia, tetany and dyspnoea during a in stimulated leukapheresis procedure on spectra optia.The patient without known allergies, risk factors or pre-existing illness had this reaction from acd-a.Per follow up with the customer, the patient was given calcium and iv magnesium.The collection set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
Investigation: a disposable complaint history search was performed for this lot and found one additional report for similar issues on this lot, also located at robert-bosch-krankenhaus.See mdr 1722028-2023-00258.According to therapeutic apheresis: a physician's handbook, adverse events occur during therapeutic procedures with a frequency of (b)(4).Transient hypocalcemia associated with apheresis is usually well tolerated.Symptoms often show as parasethesia (tingling) but patients may also experience unusual taste, nausea, lightheadedness, shivering, and tremors.Severe hypocalcemia may also cause muscle contractions and can progress to tetany and seizures if hypocalcemia escalates and is not corrected.The run data file (rdf) was analyzed for this event.This procedure indicated the system operated as intended and the procedure was run within standard operating limits.A review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.Root cause: a root cause assessment was performed for this complaint.A definitive root cause for the patient's reaction could not be determined.Possible causes for the alleged citrate reaction include but are not limited to ac management during the procedure, patient disease state, and/or patient sensitivity to anticoagulant.
 
Event Description
The customer reported severe citrate-related hypocalcaemia symptoms such as paresthesia, tetany and dyspnoea during a in stimulated leukapheresis procedure on spectra optia.The patient without known allergies, risk factors or pre-existing illness had this reaction from acd-a.Per follow up with the customer, the patient was given calcium and iv magnesium.The collection set is not available for return because it was discarded by the customer.Due to eu personal data protection laws, the patient information is not available from the customer.Donor gender and weight were obtained from the run data file (rdf).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRA OPTIA
Type of Device
SPECTRA OPTIA IDL SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key17447368
MDR Text Key320285482
Report Number1722028-2023-00259
Device Sequence Number1
Product Code LKN
UDI-Device Identifier05020583123205
UDI-Public05020583123205
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K183081
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 08/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number4123201
Device Lot Number22063083
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/03/2023
Initial Date FDA Received08/02/2023
Supplement Dates Manufacturer Received08/03/2023
09/08/2023
Supplement Dates FDA Received08/24/2023
09/18/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/03/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age38 YR
Patient SexMale
Patient Weight72 KG
-
-