Model Number 1944/52
Device Problems Failure to Capture (1081); Fracture (1260); Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2023
Event Type  Injury  
Event Description
It was reported, during in clinic follow up.That the atrial lead exhibited loss of capture and loss of sensing.Fluoroscopy revealed, that the lead was fractured.The lead was explanted and successfully replaced.The patient was stable and asymptomatic.
 
Manufacturer Narrative
The reported events of failure to capture, failure to sense and lead fracture were not confirmed.As received, a complete lead with stylet inserted was returned in one piece.Electrical testing and x-ray examination did not find any indication of conductor fractures or internal shorts.Visual inspection of the lead did not find any anomalies except for the damages found consistent with procedural damage.All tines were intact and undamaged.
 
Manufacturer Narrative
A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISOFLEX OPTIM LEAD
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17450502
MDR Text Key320303204
Report Number2017865-2023-37472
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734502320
UDI-Public05414734502320
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P960030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1944/52
Device Lot NumberP000151061
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/20/2023
Initial Date FDA Received08/03/2023
Supplement Dates Manufacturer Received08/31/2023
10/18/2023
Supplement Dates FDA Received09/07/2023
10/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-