• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 VIT E NEUTRAL LNR 36MM F; PROSTHETIC, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 VIT E NEUTRAL LNR 36MM F; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Device Problem Packaging Problem (3007)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/10/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).The product has been received by zimmer biomet.The investigation is currently in process.Once the investigation has been completed, a follow-up report will be submitted.
 
Event Description
It was reported that after the implant was opened, a hair was found directly on the implant.The surgeon waited for another liner to be delivered that same day.Once it arrived 20-25 mins later, the new one was implanted.They continued with the other steps of the procedure while waiting for the new liner to show up.Total wait time for the patient was 10 minutes.There is no additional information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product found the packaging has been previously opened.Debris was found in the blister; however, no hair-like fiber could be located.This complaint cannot be confirmed as the packaging was previously opened and the source of the debris cannot be confirmed.Review of the device history records identified no deviations or anomalies during manufacturing medical records were not provided.A definitive root cause or condition of the device when it left zimmer biomet cannot be determined, as the product was returned opened.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
There is no update to the prior event description provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 VIT E NEUTRAL LNR 36MM F
Type of Device
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17450908
MDR Text Key320348901
Report Number0001822565-2023-02029
Device Sequence Number1
Product Code LPH
UDI-Device Identifier00889024519916
UDI-Public(01)00889024519916(17)280502(10)65967754
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number30103606
Device Lot Number65967754
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/10/2023
Initial Date FDA Received08/03/2023
Supplement Dates Manufacturer Received08/16/2023
Supplement Dates FDA Received08/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexFemale
-
-