It was reported that a patient presented with high capture thresholds noted on the right ventricular lead.It was suspected the cardiac tissue of the patient did not conduct the pacing pulse well.During the surgical intervention, the lead helix was unable to retract, resulting in an operation issue.The rv lead was explanted and replaced on (b)(6) 2023.The patient was stable throughout.
|
The reported events were failure to capture, a helix mechanism issue, and operation issue.As received, a complete lead was returned for analysis.The reported event of a helix mechanism issue was confirmed.Visual and x-ray examination of the lead found the inner coil was over torqued at the connector pin region, helix extended and clogged with blood / tissue.After cutting, cleaning, and applying torque directly to the inner coil, the helix could be retracted and extended.The full helix extension length was measured within specification.The reported event of failure to capture was not confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.
|