• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA II DUODENOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA II DUODENOVIDEOSCOPE Back to Search Results
Model Number TJF-Q180V
Device Problems Crack (1135); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/05/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for inspection, and the customer's issue of ¿wire rope of the albarran lever (forceps elevator) is torn" was confirmed.The following findings were noted during device evaluation: broken albaric lever, the technician found due to wear of forceps elevator wire, forceps do not rise-up at all, air/water cylinder has no color, suction-cylinder has no color, due to deformation of electrical connector guide pin, water tightness is lost, due to a dent on plastic distal end cover, water tightness is lost, due to the wear of angle wire, bending angle in up direction does not meet specifications, connecting tube has a wrinkle, forceps elevator has foreign material or stain, and water tightness of forceps elevator is lost.The investigation is ongoing, a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus that the evis exera ii duodenovideoscope the wire rope of the albarran lever (forceps elevator) is torn during an unknown procedure.Upon inspection and evaluation, forceps elevator with foreign material and with leak.There was no report of patient harm or user injury associated with this event.This medical device report (mdr) is being submitted to capture the reportable malfunction found during incoming inspection and evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.The event likely occurred due to stress such as impact caused by equipment handling.The detection method is described in the operation manual, chapter 3 preparation and inspection and prevention measures are described in the reprocessing manual chapter 5 reprocessing the endoscope.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17456992
MDR Text Key320822186
Report Number9610595-2023-11154
Device Sequence Number1
Product Code FDT
UDI-Device Identifier04953170367311
UDI-Public04953170367311
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K143153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberTJF-Q180V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/06/2023
Initial Date FDA Received08/04/2023
Supplement Dates Manufacturer Received08/18/2023
Supplement Dates FDA Received08/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/12/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-