The device was returned to olympus for inspection, and the customer's issue of ¿wire rope of the albarran lever (forceps elevator) is torn" was confirmed.The following findings were noted during device evaluation: broken albaric lever, the technician found due to wear of forceps elevator wire, forceps do not rise-up at all, air/water cylinder has no color, suction-cylinder has no color, due to deformation of electrical connector guide pin, water tightness is lost, due to a dent on plastic distal end cover, water tightness is lost, due to the wear of angle wire, bending angle in up direction does not meet specifications, connecting tube has a wrinkle, forceps elevator has foreign material or stain, and water tightness of forceps elevator is lost.The investigation is ongoing, a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.The event likely occurred due to stress such as impact caused by equipment handling.The detection method is described in the operation manual, chapter 3 preparation and inspection and prevention measures are described in the reprocessing manual chapter 5 reprocessing the endoscope.Olympus will continue to monitor field performance for this device.
|