ICU MEDICAL COSTA RICA LTD. PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 146870489 |
Device Problem
Material Rupture (1546)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/09/2023 |
Event Type
malfunction
|
Event Description
|
The event involved a primary plum set, clave secondary port, clave y-site, secure lock, 103 inch.The reporter stated, ¿ruptured at the connector¿.No harm was reported.
|
|
Manufacturer Narrative
|
The device is not available to be returned for evaluation; however, a photo was provided and it is awaiting an investigation.
|
|
Manufacturer Narrative
|
A photo was returned showing a bag spike broken at the insert of a semi rigid bottle.No samples were returned for investigation.There was plastic deformation observed at the break.The reported complaint of a broken piercing pin can be confirmed from the photo provided.The break with plastic deformation is typical of an unintentional bending force during use.The device history record could not be reviewed due to the unknown lot number.
|
|