Catalog Number RJA108 |
Device Problems
Material Separation (1562); Device Fell (4014)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Event Description
|
According to available information, this device required replacement due to disconnection.When the metal stylet was withdrawn, the connector came off.The metal stylet could not be withdrawn from device.The device and metal stylet was removed and replaced.No other adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
|
|
Event Description
|
According to the available information, the white luer came off on a kolibri product.
|
|
Manufacturer Narrative
|
The review of the complaint history database, revealed one similar incident for the lot number 9054015.Checking the quality database revealed one non-conformity related to this issue.In august we received one used sample.According to the observation, white connector was unglued, no more issue or damage was observed.
|
|