• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN ORANGEMED, INC. NIHON KOHDEN NKV-330 VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN ORANGEMED, INC. NIHON KOHDEN NKV-330 VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number NKV-330
Device Problems Defective Component (2292); Low Readings (2460)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/10/2023
Event Type  malfunction  
Manufacturer Narrative
Nihon kohden orangemed inc.Will submit a supplemental report if additional information becomes available.
 
Event Description
It was reported that the nkv-330 ventilator kept alarming ''low fio2'' during patient use.The set fio2 was 40% and the measured fio2 was 27-34%.The patient was not harmed with spo2 and breathing rate normal.Patient was removed from this ventilator and placed on another ventilator.
 
Manufacturer Narrative
Evaluation completed.See attached failure investigation summary report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIHON KOHDEN NKV-330 VENTILATOR SYSTEM
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
NIHON KOHDEN ORANGEMED, INC.
1800 e. wilshire ave.
santa ana CA 92705
Manufacturer (Section G)
NIHON KOHEN ORANGEMED, INC.
1800 e. wilshire ave.
santa ana CA 92705
Manufacturer Contact
sheryl higgins
1800 e. wilshire ave.
santa ana, CA 92705
949502644
MDR Report Key17470389
MDR Text Key320503619
Report Number3014631252-2023-00022
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00843685100043
UDI-Public00843685100043
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K213521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 01/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNKV-330
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/07/2023
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/13/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age80 YR
Patient SexMale
Patient Weight129 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-