The device was received for evaluation.During functional testing, the reported 'did not restart after downstream occlusion' was confirmed.A service history review was performed and revealed that the device has no previous service events; therefore, servicing did not cause or contribute to the reported event.The reported condition was verified.The cause of the condition was determined to be the out of specification downstream sensor.The force sensor requires calibration to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
|