• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLFLEX BILIARY; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLFLEX BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00570740
Device Problems Break (1069); Activation, Positioning or Separation Problem (2906); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/18/2023
Event Type  malfunction  
Manufacturer Narrative
Block d4, h4: the complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.Block h6: imdrf device code a0501 captures the reportable event of inner sheath detached.
 
Event Description
It was reported to boston scientific corporation on (b)(6) 2023, that a wallflex biliary rx partially covered stent was to be implanted in the common bile duct to treat a malignant biliary stricture during an endoscopic retrograde cholangiopancreatography (ercp) with stent placement procedure performed on (b)(6) 2023.The patient's anatomy was tortuous and was not dilated prior to stent placement.During the procedure, the handle broke, the inner sheath snapped inside the catheter, and the stent could not be fully deployed.The stent was removed from the patient partially deployed on the delivery system.The procedure was completed with another wallflex biliary stent.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX BILIARY
Type of Device
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17474747
MDR Text Key320625237
Report Number3005099803-2023-04218
Device Sequence Number1
Product Code FGE
UDI-Device Identifier08714729759331
UDI-Public08714729759331
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140630
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00570740
Device Catalogue Number7074
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/21/2023
Initial Date FDA Received08/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-