ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
|
Back to Search Results |
|
Catalog Number B12LT |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/01/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during a gastric bypass, a gauze pad was put in and when it was removed, the inner gasket of the sleeve pulled out.The rep was present during the case and it was stated that the gauze pad was removed incorrectly.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 8/7/2023 b3: unknown, assumed first day of month that complaint was reported d4: batch # unk a manufacturing record evaluation is pending.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 8/30/2023.D4: batch # x7016u.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the b12lt device was received with the duckbill out of position and present.The device was visually inspected and no damaged found on the sleeve.No conclusion could be reached as to what might have caused the duckbill out of position.As part of ethicon¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
|
|
Search Alerts/Recalls
|
|
|