SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD TEXIUM¿ NEEDLE-FREE SYRINGE; PISTON SYRINGE Back to Search Results
Catalog Number MY8030-0006
Device Problems Leak/Splash (1354); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/19/2023
Event Type  malfunction  
Event Description
It was reported that a disconnection occurred between the bd texium¿ needle-free syringe and texium connector, causing chemotherapy medication to leak into the isolator's working chamber.This complaint was created to capture the 2nd of 2 related incidents.The following information was provided by the initial reporter, translated from italian: "disconnection between texium connector and syringe resulting in drug spillage in the weeks from (b)(6) 2023 to (b)(6) 2023.Five incidents occurred with texium syringes code my8030-0006 from the company becton dickinson batch no.92220701 with expiry date 26.07.2025.Specifically, when the chemotherapy drug was withdrawn from the bottle, a disconnection occurred between the texium connector and the syringe with consequent spillage of chemotherapy drug (irinotecan and 5-fu).It should be noted that these incidents occurred within the isolator's working chamber, thus avoiding contamination of the operator and patient".
 
Manufacturer Narrative
B.3.Date of event: unknown.The date received by manufacturer has been used for this field.H.3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Manufacturer Narrative
The following fields have been updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 27-sep-2023.H.6.Investigation summary: two hundred and fifty samples from lot 92220701 were received in sealed packaging for investigation.The feedback provided by the customer suggests a separation of the texium was observed from the syringe resulting in leakage.A visual inspection of the returned samples did not identify any obvious damage or manufacturing defects which could have caused or contributed to the customer's experience.Twenty-five samples were then subjected to functional testing by manually applying force at the texium connection; no separation was detected throughout testing.The samples were then connected to a bd maxzero device from stock; again a secure connection was detected and no flow issues were detected throughout testing.The samples were then subjected to pressure testing; no leakage was detected at the texium connection throughout testing.The details of this feedback were forwarded to the manufacturing site for investigation.In this instance, a definitive root cause could not be identified as testing of the returned samples did not identify any product defects that could have contributed to the customer¿s experience.The alleged complaint sample was not available for investigation therefore it was not possible to confirm whether a manufacturing defect may have contributed to the customer's experience.A review of the production records for lot 92220701 did not identify any in-process testing failures or quality deviations which may have resulted in a report of this nature.The quality team at the manufacturing site has been informed of this complaint in order to be aware of the reported failure mode during future production of this product.A review of the customer feedback database indicates that this is an isolated feedback with no further reports of this nature against the my8030-0006 product over the past 12 months.H3 other text : see h10.
 
Event Description
It was reported that a disconnection occurred between the bd texium¿ needle-free syringe and texium connector, causing chemotherapy medication to leak into the isolator's working chamber.This complaint was created to capture the 2nd of 2 related incidents.The following information was provided by the initial reporter, translated from italian: "disconnection between texium connector and syringe resulting in drug spillage in the weeks from 03.07.2023 to 18.07.2023.Five incidents occurred with texium syringes code my8030-0006 from the company becton dickinson batch no.92220701 with expiry date 26.07.2025.Specifically, when the chemotherapy drug was withdrawn from the bottle, a disconnection occurred between the texium connector and the syringe with consequent spillage of chemotherapy drug (irinotecan and 5-fu).It should be noted that these incidents occurred within the isolator's working chamber, thus avoiding contamination of the operator and patient".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD TEXIUM¿ NEEDLE-FREE SYRINGE
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
jennifer suh
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key17477456
MDR Text Key320873525
Report Number9616066-2023-01670
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberMY8030-0006
Device Lot Number92220701
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/19/2023
Initial Date FDA Received08/07/2023
Supplement Dates Manufacturer Received09/29/2023
Supplement Dates FDA Received10/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-